Real time PCR equipment

The Center bases its work on the following instruments:

The available applications:

Consumables and Disposables

  • ABI PrismTM 96-well Optical Reaction Plates sealed with Optical Adhesive Covers
  • ABI PrismTM 384-well Optical Reaction Plates sealed with Optical Adhesive Covers
  • Micro Fluidic Cards

Real-Time PCR using the SYBR Green or TaqMAN:

There are three options:

  1. Running a 96-wells PCR plate, the assay is done by the customer.
  2. RNA samples are submitted. RT-PCR amplifications and Real-Time PCR are performed at the Center, with primers provided by the customer.
  3. cDNA samples are submitted. Real-Time PCR is performed at the Center, with primers provided by the customer.

Requirements for Primers (SYBR-GREEN chemistry)

  • From 30 to 80% G/C content.
  • Melting temperature (Tm) from 58 to 60°C.
  • The five nucleotides at the 3´ end of primer should have no more than two G or C bases.
  • Recommended primer concentration 50nM.  
  • Length of amplicons should be in 50-150 basepairs range.

Requirements for Primers and Probe (TaqMAN chemistry)

  • Primers and probe with 30 to 80% G/C content
  • Melting temperature (Tm) for the primers 58 to 60°C, Tm for the probe 8 to 10°C higher.
  • Do not select probes with G on the 5´ end.
  • The five nucleotides at the 3´ end of primer should have no more than two G or C bases.
  • Recommended primer concentration 900nM and probe concentration 250nM.
  • Amplicons are in the 50- to 150-basepairs range.   

SNP (Single Nucleotide Polymorphism) detection using the TaqMAN Probe

SNP Assays provided by the customer.

Genomic DNA samples are submitted

The DNA concentration should be 10-20 ng/µl.

The volume of the samples depends on the number of requested SNPs but not less than 20 µl.

DNA samples should arrive in a 96-wells plate even if the plate is not full.

The order of the samples in the plate should always be as follows:

          Sample 1 is in position A1

          Sample 2 is in position B1

          .....................................

          Sample 96 is in position H12

High-throughput (96x96) RT PCR on Fluidigm BioMark.

Real-Time PCR is performed at the Center.

IFC-array provided by the customer.

For Gene Expression analysis the customer need to provide plate with cDNA samples and plate with diluted primers. The final concentration of Forward+Reverse primers should be 50µM.